Skip to content

CP 104 X-Ray Generator

The CP Series has continued to evolve over its product life and is available in the range of 100 to 450 kV with both unipolar and bipolar options. CP series generators are available in single or dual focus configurations. The CP series uses 25 kHz for EHT generation and the family of products offers low ripple and low drift.
  • All units connect to the x-raytube with industry standard rubber coned connectors designated R10, R24 or R30.
  • All units are capable of operation with external and embedded controllers.
  • Oil is used as an insulation medium together with epoxy encapsulated sub assemblies. No vacuum impregnation is required and local servicing is possible.

Technical Specification

kV Range 10 - 100mA Range 0 – 75 within a 3000 W limit
Line Regulation < 0.2% change for an input change of +/- 10% in 2 SecondsLoad Regulation < 0.2% change for a load change from 100uA to 30mA in 5 Seconds
Output Drift 0.1% over 8 hours after initial 30 minute equalisationOutput Risetime 20kV per second.
Output Accuracy +/- 2.5%Filament Supply Up to 9V at 4.5A
Output Connector R24Environmental Conditions 10 to 35 Deg. C at up to 80% Humidity, non condensing
Input Voltage 220 V AC +/- 10%
47 – 63 Hz
Multiplier Frequency 25 KHz
Size mm (WxLxH) 435 x 667 x 729Input Current 27A at 220V AC
Weight 220